ISSN 1016-5169 | E-ISSN 1308-4488
pdf
Treatment of severe hemolysis following Nit-Occlud Lê VSD coil implantation with Amplatzer Duct Occluder II [Turk Kardiyol Dern Ars]
Turk Kardiyol Dern Ars. 2016; 44(7): 593-596 | DOI: 10.5543/tkda.2016.69841

Treatment of severe hemolysis following Nit-Occlud Lê VSD coil implantation with Amplatzer Duct Occluder II

Mustafa Orhan Bulut, Mehmet Küçük, Şevket Ballı, Ahmet Çelebi
Siyami Ersek Education And Research Hospital, Pediatric Cardiology, Istanbıl

Hemolysis after percutaneous device closure of ventricular septal defect (VSD) is a rare complication that can be conservatively managed in the majority of cases. However, surgery or transcatheter occlusion may be necessary. Presently described is the successful treatment of hemolysis that developed after implantation of the Nit-Occlud Lê VSD coil, using the Amplatzer Duct Occluder II (ADO II) in a patient with aneurysmatic perimembranous VSD. Systolic murmur and symptoms immediately disappeared after the procedure.

Keywords: Coil implantation, hemolysis, transcatheter closure, ventricular septal defect.

Corresponding Author: Şevket Ballı, Türkiye
Manuscript Language: English
×
APA
NLM
AMA
MLA
Chicago
Copied!
CITE


Journal Metrics

Journal Citation Indicator: 0.18
CiteScore: 1.1
Source Normalized Impact
per Paper:
0.22
SCImago Journal Rank: 0.348

Quick Search

Copyright © 2025 Archives of the Turkish Society of Cardiology



Kare Publishing is a subsidiary of Kare Media.